Beurer EM 89 Heat Instrucciones De Uso página 35

Electroestimulador tens/ems con función de calor

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 59
/
4. Use the
adjustment buttons to select your preferred treatment time
and press the ENTER button to confirm.
/
5. Use the
adjustment buttons as relevant in each case for
and
to select your preferred pulse intensity.
Ch3
Ch4
EMS 33 program
EMS 33 is a program that you can also customise. In this program, you can set
the pulse frequency to between 1 and 150 Hz and the pulse width to between
80 and 320 µs.
1. Position the electrodes on your preferred treatment area (for positioning
suggestions, see "Electrode positioning") and connect them to the device.
2. Select the EMS 33 program as described in the "Starting use" section (step
3 to step 5).
/
3. Use the
adjustment buttons to select your preferred pulse frequency
and press the ENTER button to confirm.
/
4. Use the
adjustment buttons to select your preferred pulse width and
press the ENTER button to confirm.
/
5. Use the
adjustment buttons to select your preferred treatment time
and press the ENTER button to confirm.
/
6. Use the
adjustment buttons as relevant in each case for
and
to select your preferred pulse intensity.
Ch4
Ch3
EMS 34 program
EMS 34 is a program that you can also customise. In this program, you can set
the pulse frequency to between 1 and 150 Hz and the pulse width to between
80 and 450 μs. You can also set the working time and resting pause for this
program to between 1 and 30 seconds each.
1. Position the electrodes on your preferred treatment area (for positioning
suggestions, see "Electrode positioning") and connect them to the device.
2. Select the EMS 34 program as described in the "Starting use" section (step
3 to step 5).
/
3. Use the
adjustment buttons to select your preferred working duration
("on time") and press the ENTER button to confirm.
/
4. Use the
adjustment buttons to select your preferred resting duration
("off time") and press the ENTER button to confirm.
/
5. Use the
adjustment buttons to select your preferred pulse frequency
and press the ENTER button to confirm.
/
/
/
6. Use the
adjustment buttons to select your preferred pulse width and
press the ENTER button to confirm.
/
7. Use the
adjustment buttons to select your preferred treatment time
and press the ENTER button to confirm.
/
8. Use the
adjustment buttons as relevant in each case for
and
to select your preferred pulse intensity.
Ch4
Ch3
EMS 35 program
The EMS 35 program is a burst program that you can also customise. In this
program, the intensity varies over time. Burst programs are suitable for all areas
of application to be treated with changing signal patterns (to minimise the level
of becoming accustomed to the treatment). In this program, you can set the
pulse frequency to between 1 and 150 Hz and the pulse width to between 80
and 450 µs. You can also set the working time and pause time for this program
to between 1 and 30 seconds each.
1. Position the electrodes on your preferred treatment area (for positioning
suggestions, see "Electrode positioning") and connect them to the device.
2. Select the EMS 35 program as described in the "Starting use" section (step
/
/
3 to step 5).
/
3. Use the
adjustment buttons to select your preferred working duration
("on time") and press the ENTER button to confirm.
/
4. Use the
adjustment buttons to select your preferred resting duration
("off time") and press the ENTER button to confirm.
/
5. Use the
adjustment buttons to select your preferred pulse frequency
and press the ENTER button to confirm.
/
6. Use the
adjustment buttons to select your preferred pulse width and
press the ENTER button to confirm.
/
7. Use the
adjustment buttons to select your preferred treatment time
and press the ENTER button to confirm.
/
8. Use the
adjustment buttons as relevant in each case for
and
to select your preferred pulse intensity.
Ch3
Ch4
35
/
/
/
/