EN-15
8.
Press the Continue key to continue high range turbidity calibration. If high range
turbidity calibration not required, press any keys to exit. (Figure 7)
9.
Fill the 10 ml sample vial to above 10 ml mark with the 100 or 200 NTU
standard. Insert the sample vial to the sample vial compartment.
10.
Follow the message prompts, use the upper or down key to switch the standard
between 100 NTU and 200 NTU.
11.
Press the Continue key to measure the selected standard. High range
turbidity calibration is successful. (Figure 8)
12.
Press any keys to exit.
2022/07/19 08:00
Figure 7 Turbidity calibration for low range
2022/07/19 08:00
Figure 8 Turbidity calibration for high range
a-AP50MM