EN-22
CONFG button to launch the method configuration page.
3.
Press the slope calibration button SlpCal. A numeric keyboard will be
displayed.
4.
Enter the concentration value and press the OK key on the enter key in the
numeric keyboard to return to the configuration page.
5.
Press the EXIT button. Press the OK key to accept the calibration or other key to
cancel the calibration.
For best results, the concentration of the standard solution should be less than the
maximum concentration for the method (table 2) and greater than the half of the
maximum concentration. For example, to calibrate total chlorine, the chlorine
concentration in the standard solution should be between 1.1 and 2.2 ppm.
The corresponding calibration parameters will be updated and saved in the memory
as the working calibration parameter set. Note that this set of calibration parameters
are not the same as the default set. You can use Default button to copy the default
calibration parameters to the working set.
2022/07/19 08:00
Figure 15. Slope Calibration