Idiomas disponibles

EN-23
5.8.3. Reagent Blank Calibration
Some methods have a non-zero intercept value in the calibration equation. For these
methods, a proper non-zero intercept value is pre-loaded in the AP50MM prior to
shipping. The following steps are used to carry out a reagent blank calibration:
1.
Follow the normal steps to carry out a measurement on a deionized water
sample.
2.
Press the CONFG button to launch the method configuration page.
3.
Press the reagent blank calibration button BLKC
4.
Press the OK key to save when exiting from the configuration page or press
other keys to cancel.
5.8.4. Resume to Default Calibration Parameters
Pressing the Default button will copy the default calibration intercept and slope to the
working intercept and slope, respectively. If the default calibration parameters were
created prior to shipping, this button action is to restore the working calibration
parameters to the original factory loaded calibration parameters.