Service Life; Warranty Information - Invacare Roze Serie Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 137
Invacare® Roze™

1.4 Service Life

WARNING!
Risk of injury
Maintenance MUST be performed only by qualified
personnel.
Improper assembly may cause injury or damage.
– Regular maintenance of patient lifts and accessories is
necessary to assure proper operation.
– DO NOT overtighten the mounting hardware. This
will damage the mounting bracket.
The stand assist lift has an expected lifetime of 8 years when used
in accordance with safety instructions, maintenance intervals and
correct use, stated in this manual. The effective product lifetime can
vary according to frequency and intensity of use.
Actuator Service Life
Number of lifts per day
1–2
3
4
5
6
7–9
10–13
14–27
8
Service life of actuator
(in years)
10
9
6
5
4
3
2
1

1.5 Warranty Information

Terms and conditions of the warranty are part of the general terms
and conditions particular to the individual countries in which this
product is sold.
Contact information for your local Invacare office is located inside
the back cover of this manual.
1164940-D

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido