Fluid Monitoring Unit Set-Up - Hologic Aquilex Instrucciones De Uso

Ocultar thumbs Ver también para Aquilex:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 179
System Operation
en
Fig. 5-3
Fluid Monitoring Unit (cart
with scale)
Bag scale
(1)
Fluid bags
(2)
Bag deflector
(3)
(4)
Container scale cable/connector
Container scale
(5)
Holder for upper container (Serres,
(6)
Medela)
Holder for lower container
(7)
(Abbott, Bemis®, Medi-Vac®,
DeRoyal®)
Roller wheel base
(8)
Container
(9)
16
CAUTION!
Do not use the covered power output at the rear of the irrigation pump unit.
WARNING!
Additional equipment
Additional equipment connected to medical electrical devices must be demon-
strated to be compliant with their respective IEC or ISO standards (IEC 60601-1,
IEC 60950 or IEC 62368 for data processing equipment).
5.3

Fluid Monitoring Unit Set-up

(1)
(9)
The fluid monitoring unit (cart/scale) consists of a weighing unit for fluid bags
(1), a weighing unit for fluid containers (5), and a roller wheel base (8).
1. Remove the cart/scale from the cardboard shipping box.
2. Remove the pump and the power cords from the first cardboard box.
3. Loosen the handwheel (6) (Fig. 5-4, page 17) and pull bag scale upwards to
stop. The screw (7) (Fig. 5-4, page 17) must be inserted into the provided
opening. Secure the bag scale with the handwheel.
4. Depending on the type of container used, attach the container rings (included
in the second box) to the upper (6) or lower (7) container holders ( Fig. 5-
3, page 16).
5. Guide the power cord through the holes provided for this purpose and con-
nect to pump (2) (Fig. 5-4, page 17) and plug into grounded, shockproof safe-
ty wall socket.
6. Attach the scale to the pump by connecting the connector of the bag scale
with the connector of the container scale (4) (Fig. 5-4, page 17) and fix the
connected cables below the lower pump tray by means of the provided wire
clips. Connect the second connector of the bag scale to the port on the back
of the pump (5) (Fig. 5-4, page 17).
(2)
(3)
(4)
(5)
(6)