System Operation
Fig. 5-6
High vacuum tube
en
Hygiene filter
(6)
Vacuum tube (green connectors)
(7)
Connection for high vacuum
(8)
(green)
(9)
Container
Tissue trap
(10)
Adjustment dial
(11)
Connecting tube
(12)
20
(11)
(6)
on ly
for
M yo
Su re
®
(7)
5.4
Turning On the Aquilex System
1. Press the ON/OFF switch. The displays and indicators light up and system
turns on.
2. The system now performs a device self-test.
The device self-test is used by the system to check whether the container and
the bag scale are connected, among others. If the bag scale is not connected,
the message "System OK. Bag Scale Not Connected" is displayed. In this case,
please check the appropriate data connection of the scales with the pump ac-
cording to chapter 5.3.
3. If a tube set is in the inflow tube holder when the pump is switched on, the
pump display (Fig. 5-1, Front of Irrigation Pump Unit (1)) shows the message
Remove Tube Set. The device self-test resumes once the tube set is removed
from the roller wheel.
If the device self-test is unsuccessful, the corresponding error messages are
displayed (see Chapter 10, Error and Warning Messages).
The system has successfully completed the device self-test when a single audible
beep is heard. The message System OK is displayed for 5 secs followed by the
message Insert Tube Set.
(12)
(8)
(10)
(9)