BTL5-A/C/E/G1_-M_ _ _ _-B/Z-NEX-S32/KA_ _
Micropulse Linear Transducer - Rod Style
9
Manual adjust mode
Enable buttons:
1. Press one of the buttons for at
least 3 s. Release button.
2. Within 1 s hold down buttons 1+2
simultaneously for at least 3 s. Now
the Error value remains stored as
the output signal.
Select adjust:
Hold down button 2 for at least 2 s
until the code for manual adjust is
displayed. Release button. The cur-
rent position value is displayed.
Adjust start value:
1. Move magnet to exact start posi-
tion.
2. Hold down button 1 for at least 2 s.
3. Shift null point (=start value) to-
wards flange or rod end with con-
stant slope: Briefly pressing the
buttons increases or decreases the
actual value by approx. 1 mV or
2 µA. Holding a button down for
longer than 1 s increases the step
size.
4. Exit calibration procedure for the
start value: Press buttons 1+2 less
than 2 s.
Adjust end value:
5. Move magnet to exact end posi-
tion.
6. Hold down button 2 for at least 2 s.
7. Increase or reduce slope: Briefly
pressing the buttons increases or
decreases the actual value by
approx. 1 mV or 2 µA. Holding a
button down for longer than 1 s
increases the step size.
8. Exit calibration procedure for the
end value: Press buttons 1+2 less
than 2 s.
Additional notes:
Setting the end value (i.e. the slope)
and the start value can mutually
affect each other depending on the
stroke position. You will need to
repeat steps 1 to 8 over until the
start and end values agree with their
respective desired values.
End adjustment, disable buttons:
Hold down buttons 1+2 simulta-
neously for at least 6 s to end the
calibration procedure until the Error
value is displayed. Then press one
of the buttons briefly (<1 s) to dis-
able both buttons.
Check your settings carefully be-
fore starting up the system.
BTL5-A... with magnet in stroke range
>3 s
1
/ 2
<1 s
>3 s
1
+ 2
>2 s
2
adjust
>2 s
1
++
1
1
+
>1 s
>1 s
–
2
– –
2
<2 s
1
+ 2
adjust
>2 s
2
++
1
1
+
>1 s
>1 s
–
2
– –
2
<2 s
1
+ 2
adjust
1
+ 2
>6 s
<1 s
1
/ 2
Ignition Class "n"
Displayed values (example)
for 0...10 V
for 4...20 mA
5.39
9.15
V
A
m
5.39
9.15
V
A
m
10.50
3.60
A
V
m
10.50
3.60
V
A
m
2.00
6.00
V
A
m
5.39
9.15
V
A
m
1.04
4.82
A
V
m
0.00
4.00
V
A
m
1.34
4.82
V
A
m
1.00
4.40
V
A
m
2.00
6.00
V
A
m
1.00
4.40
A
V
m
9.89
19.13
V
A
m
10.00
20.00
V
A
m
9.89
19.13
V
A
m
9.00
19.00
V
A
m
2.00
6.00
V
A
m
9.00
19.00
V
A
m
10.50
3.60
V
A
m
10.50
3.60
V
A
m
9.00
19.00
A
V
m
Current position
value
Current position
value
Code = Error value
Code = Error value
Code for adjustment
Current position
value
Current position
value
Code for adjusting
start value = Null value
Current position
value
New start value
Code for adjustment
Current position
value
Current position
value
Code for adjusting
end value = End value
Current position
value
New end value
Code for adjustment
Current position
value
Code = Error value
Code = Error value
Current position
value
11
english