Warning Signs; Error Signs - ACTEON QUAZ 17 l Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 45
Vacuum Test printing format
Label printing format

5.4. WARNING SIGNS

As well as the information that appears on the screen during cycles and the error messages,
the QUAZ autoclave has a series of warning signs that appear on the screen so the user can
act accordingly.

5.5. ERROR SIGNS

The error signs may appear while the autoclave is performing any of the stages or during the constant analysis that the machine
carries out of the parameters of the process.
In all of these cases the autoclave will come to a full stop, there will be a series of short beeps and the error information will remain
on the screen until the user presses reset. The autoclave does not unlock the door until the sterilisation chamber reaches an
atmospheric pressure of ± 60 Mbars.
The error signs that the QUAZ autoclave displays are the following:
Filling time exceeded.
ERROR 001
The distilled water tank takes longer to fill up than the established time.
Not enough heat.
ERROR 002
- One of the heating devices it not working properly.
(020,021,022,
- The temperature has not been restored within the established time.
023)
- Sterilisation has not begun within the established time.
Excessive pressure.
ERROR 003
The pressure inside the sterilisation chamber has exceeded the established
limits for the stage in process.
Low pressure.
- The pressure has not been restored within the established time.
ERROR 004
- The machine has not reached the sterilisation pressure within the established
(040,041,042)
time.
- In the stages when the pressure increases, the pressure has not increased in
the established way.
Cycle printing format
DESCRIPTION
Maintenance warnings:
Information warnings:
The warning screens that are described below may appear when a cycle is NOT in process.
- Excessive temperature (when a cycle begins).
- Low power supply voltage.
- Filling time exceeded.
- Pressurised chamber (if the user tries to open the door
with higher or lower pressure).
- SD card not accessible.
- Data not saved. Check SD.
- Water quality not acceptable.
- Excessive pressure (on the idle screen).
- Excessive temperature (on the idle screen).
- Temperature system failure.
- Pressure system failure.
- Power supply failure.
- Level detector failure.
- Door malfunction.
-System Error 01, 02.
Make sure that the union joints are connected properly and that there is enough
supply of distilled water.
Check that the distilled water filter is not obstructed.
If the problem persists, please contact the Technical Support Dept.
Make sure that the power supply is 230 + 10%.
Allow the autoclave to cool down and repeat the cycle.
If the problem persists, please contact the Technical Support Dept.
Allow the autoclave to cool down and repeat the cycle.
If the problem persists, please contact the Technical Support Dept.
Make sure that the power supply is 230 + 10%.
Allow the autoclave to cool down and repeat the cycle.
If the problem persists, please contact the Technical Support Dept.
ACTION
17

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Quaz 24 l

Tabla de contenido