2.5 Explanation of symbols (connections and nameplate)
alternating current
for serial numbers < 20,000:
protection class i equipment. The medical device must be
connected to a system with protective earth conductor!
for serial numbers > 20,000:
protection class ii equipment. The medical device has a
double or reinforced insulation.
type b applied part
power switch OFF
power switch On
The number next to this factory symbol is the year of manufacture
The name next to this factory symbol is the manufacturer
The number next to this symbol is the article reference number
device complies with Council directive 93/42/EEC about medical devices,
tested and approved by dQs Medizinprodukte Gmbh
The number next to this symbol is the serial number
m
Caution! Observe warnings set forth in operation manual!
Observe Operation Manual
do not dispose with unsorted municipal waste
Protect from moisture
56
IP21
The iP rating indicates the level of protection and thus the suitability of the
device for use under diferent ambient conditions.
The rating iP21 means:
2 is the level of protection against contact and solid objects
The digit 2 means:
– Protection from contact: protected from contact with a inger
– Protection against foreign object: protected against solid foreign object
(diameter of 12.5 mm and greater)
1 indicates the degree of protection against water
The digit 1 means: protection against vertically falling water drops
ARTROMOT®-E2
do not rest or lean on or push the assembled device because it might
tip over.
To prevent tip-over of the assembled ARTROMOT®-E2 during transport,
m
choose the following settings:
1 a = 0° / 1 b = 270° / 1 c = 0° / 2 = 3 / 3 = 90° / 4 = 0 / 5 = 0
ARTROMOT®-E2 compact
To prevent tip-over of the assembled ARTROMOT®-E2 compact during
m
transport, choose the following settings:
1 = nicht zutrefend / 2 = 0 / 3 = 90° / 4 = 0 / 5 = 0