Guidance and manufacturer's declaration – electromagnetic immunity
This breast pump is intended for use in the electromagnetic environment specified below. The customer
or the user of the breast pump should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
outside ISM band
6 Vrms
150 kHz to 80 MHz
in ISM and amateur
radio bands
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Compliance
Electromagnetic
level
environment – guidance
Except as indicated in the table on the
next page, portable and mobile RF
communications equipment should be used
no closer to any part of the breast pump,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommend separation distance
d
3 Vrms
= 1.2
d
6 Vrms
= 2.0
d
10 V/m
= 0.35√P 80 MHz to 800 MHz
d
= 0.7√P 800 MHz to 2.5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol.
P
P
should be less than the compliance
b
53