5.7. vARIATIONS IN OPERATING RANGE

Range of
Influential quantity
influence
Battery voltage
9 V - 12 V
Temperature
-10°C +55°C
Humidity
20% - 80% HR
Frequency
15 - 500 Hz
AC voltage
superimposed on the
0% Un - 20% Un
Test voltage
4: The terms DAR and PI and the capacitance and current leak measurements are included in the
quantity "MΩ".
quantity
influenced
4
V
MΩ
V
0.15% /10°C
MΩ
0.20% /10°C
V
MΩ (10 kΩ to 40 GΩ)
MΩ (40 GΩ to 10 TΩ)
V
MΩ
0.1% / % Un

Influence
Typical
Maximum
< 1 pt
2 pt
< 1 pt
3 pt
0.3% /10°C +1 pt
1% /10°C + 2 pt
0.2%
1% +2 pt
0.2%
1% +5 pt
0.3%
15% +5 pt
3%
0.5% +1 pt
0.5% / % Un +5 pt