12K19=40
The friction (restraint) setting of the internal/external rotational movement
(sickle movement) in the elbow set-up can be changed by tightening the
adjustment screws (Fig. 2, item A).
12K5=* / 12K20=*
The friction (restraint) setting of the rotational movement (sickle movement)
in the elbow set-up can be changed by tightening the 13G11/ 13G39 Set
Screw (Fig. 3).
For myoelectric arm prostheses, the optional 9E185=50-1 extension cable
can be used and routed through the elbow joint.
3 Technical data
Article number
Weight (g)
Upper arm connection
⌀ (mm)
Wrist connection
⌀ (mm)
4 Handling
NOTICE
Incorrect alignment, assembly or adjustment
Product damage due to incorrectly installed or adjusted prosthetic com-
ponents
Assembly, adjustment and maintenance operations may only be
f
completed by qualified personnel.
Observe the alignment, assembly and adjustment instructions.
f
NOTICE
Mechanical damage to the product
Change in or loss of functionality due to damage
16 | Ottobock
12K19=40
260
54
40
12K5=45
12K20=45
370
67
45
12K5=50
12K20=50
330
50