Service Life - Ottobock AeroLink 6A50 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 52
2 Application
2.1 Intended Use
The product is to be used solely for lower limb prosthetic fittings.
2.2 Area of Application
Approved for a body weight of up to 125 kg.
The product must be used in TT prostheses only.
2.3 Environmental Conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)

2.4 Service Life

This prosthetic component has been tested by the manufacturer for 3 million
load cycles according to ISO 10328. Depending on the patient's activity
level, this corresponds to a service life of 3 to 5 years.
3 Safety
3.1 Explanation of Warning Symbols
3.2 General Safety Instructions
CAUTION
Using the product without following the instructions for use
Deteriorating health condition and damage to the product because of fail­
ure to follow the safety information
► Observe the safety information in these instructions for use.
► Make the patient aware of all safety instructions marked with "Inform
the patient".
CAUTION
Excessive strain on the product
Fall due to breakage of load-bearing components
Ottobock | 19