CAUTION!
Drying time depends on several variables, including: altitude, humidity, type of
wrap, preconditioning, size of chamber, mass of load, and placement in cham-
ber. Users must verify that drying time set in their autoclave yields dry surgical
equipment.
CAUTION!
Gamma Ray Sterilization
Do not perform a gamma ray sterilization.
*Note
The manufacturer is liable for product defects for the duration of three years in-
sofar they already existed at the time of the risk transfer. Excluded are failures or
outages due to wear and tear of the manual remote control causes by sterilizing
the unit. However, if the user is able to provide evidence that the number of ster-
ilization cycles amounted to less than 100 at the time of failure, the product is re-
placed free of charge.
11.6
Replacing the Fuse
CAUTION!
Before replacing the fuse, check the values of the fuse to be inserted acc. to chap-
ter 13 "Technical Data".
The fuse may be defective and is in need of replacement if:
• displays and LEDs (if available on your equipment) do not light up,
• the device does not function.
Check to make sure
• the main power supply cable is properly connected to the power supply input
and to a safety socket,
• the house power supply fuse is functioning.
WARNING!
Unplug the power cable from the device before checking the fuse.
The device does not have to be opened to replace the fuse.
1. Switch device off.
2. Disconnect device from power supply.
3. Remove power connection cable from mains socket.
4. The fuse holder is located next to the mains socket.
5. Remove fuse holder as depicted in Fig. 11-1 "Opening the fuse holder".
6. A Undo the latch of the fuse holder with a small screwdriver.
7. B Remove the fuse holder.
8. C Check fuse.
9. Insert a new fuse. Use only the specified type of fuse (see chapter 13 "Techni-
cal Data").
10. Insert the fuse holder until it can be heard snapping into place.
11. Use the power cable to reconnect the shockproof safety socket with the rear
mains socket.
Care and Maintenance
45