Stryker Flosteady Manual De Instrucciones página 19

Ocultar thumbs Ver también para Flosteady:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Make sure the connection data and technical specifications of the power supply
comply with DIN VDE or national requirements. The mains connection cable may
be plugged only into a properly installed, grounded safety wall socket (shock-
proof socket) (see DIN VDE 0100-710). Read the device label located in rear of de-
vice (type plate) to determine the operating voltage of the device.
WARNING!
To avoid risk of electrical shock, this equipment must only be connected to a sup-
ply mains with protective earth.
Only use a certified (UL-listed), removable mains connection cable, type SJT, min-
imal 18 AWG, 3 leads. The plug connectors must comply with NEMA 5-15 or IEC
60320. Grounding will only be reliable if the equipment is connected to a corre-
sponding hospital grade socket.
Medical devices are subject to special safety and protective measures concerning
electromagnetic compatibility (hereafter abbreviated as EMC).
This device is to be used only for the purposes described in the manual and has
to be installed, set up, and operated in compliance with the EMC notes and in-
structions.
Initial Device Startup
Only for U.S. operators
Precautionary measures
EN
13

Publicidad

Tabla de contenido
loading

Tabla de contenido