Safety Functions
EN
32
CAUTION!
The message "Critical Overpressure" overwrites all other possible messages.
Overpressure warnings serve as information for the attending physician to re-
spond appropriately to the existing excess pressure. The intra-articular joint
pressure can be reduced by opening the outflow cannula. If possible, the attend-
ing physician should perform this task.
When the pressure drops below 250 mm Hg, the warning Critical Overpressure
disappears and the pump resumes its normal functionality.