Error and Warning Messages

Error message
Display turns off
1 acoustic signal, cyclical
Device error/Call service/Electronics
error
2x acoustic signals, cyclical
Device error/Call service/Sensor
error
3 acoustic signals, cyclical
Device error/Call service/Motor error
5 acoustic signals, cyclical
Device error/Call service/Keyboard
error
9 acoustic signals, cyclical
Device error/Call service/ Calibra-
tion error
10 acoustic signals, cyclical
Device error/Tube detection error
Roller wheel stops
6 acoustic signals, cyclical
*These steps may only be performed by an authorized service technician.
Warning Messages
Insert tube set
3 acoustic signals cyclical
Critical overpressure
Actual pressure display flashes
1 acoustic signal, cyclical
Remove tube set
Cause
• Communication error
• Incompatible software ver-
sions on the modules (soft-
ware updates incomplete)
• Board, plug, cable defective
• Checksum error of the pro-
gram memory or an EEPROM
• Pressure sensor exceeds max.
permissible limits
• Motor blocked, defective, or
not connected
• Roller wheel defective
• Impulse transmitter defective
• Board defective
• Keyboard defective
• Keyboard not connected
• Main board defective
• Sensor not calibrated
• Tube detection defective
Cause
• Attempted device start with-
out inserted or improperly
inserted tubing
• Safety threshold for pressure
exceeded
• Device was switched on with
disposable tubing inserted,
which is the reason the sen-
sor could not be checked
Error and Warning Messages
Troubleshooting
• Check cable/plug/board and replace if necessary.*
• Reload software.*
• Check plug, cables, boards and replace if neces-
sary.*
• Reload software.*
• Replace sensor.*
• Check roller wheel, must rotate.
• Check motor connection cable.*
• Replace motor.*
• Check plug, cables, boards and replace if neces-
sary.*
• Connect/replace keyboard.*
• Replace main board.*
• Calibrate sensor.*
• Turn the device off and back on after approx. 10
seconds have expired.
If the error message is displayed again, the device
may not be used any longer. Make sure the device
can no longer be operated until a qualified service
technician conducts the appropriate tests and
repairs.
Troubleshooting
• Insert tubing correctly.
• Realign pump head with reference centering block.
• Reduce pressure. The pump starts automatically
with the preset values once the pressure has been
reduced below the critical threshold value.
• Remove tubing from pump head. Restart the device
without inserted disposable tubing.