MT10 Instruction for Use - EN
Operating Language
To set the operating language (English, French or German) use the options within the
CONFIGURATION menu
The function of the left and right keys is usually shown on the bottom line of the display. When not
performing a test the MT10 will switch off automatically if no key is pressed for 90 seconds. This time
can be extended to 180 seconds in the CONFIGURATION menu.
The LEDs indicate the status of the system:
Green LED
Off
On
Off
Slow flash
Off
On

3.3 The Probe

The small holes through the MT10 probe tip must be kept clear. If these become blocked a warning
message will be displayed. The tip must be removed and cleaned or replaced.
To remove the tip, unscrew the nose cone and pull the tip off the probe boss. A small seal will be
found in the base of the probe tip. This should be examined and replaced if it is damaged.
When replacing the tip, ensure that the seal is correctly inserted with its flat aligned with the flat on the
probe tip. Push the probe tip over the boss and replace the nose cone. Make sure that the nose cone
is screwed home firmly but do not over-tighten. Do not use any tools to tighten the nose cone.
After replacing the tip a Daily Check should be carried out.
Yellow LED
Off
MT10 turned off
Off
Idle & ready to use
Slow flash
Attempting to obtain an ear seal
Off
Taking a measurement
Fast flash
Pump error at switch on
Flickering
Sending data to a PC
Page 6