7. EMC / Technical description (cont.)
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 4)
Electromagnetic immunity
The electric breastpump Swing is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breastpump Swing should assure
that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
34
IEC 60601
Compliance
test level
level
3 Vrms
150 kHz to
80 MHz
3 Vrms
3 V/m
3 V/m
80 MHz to
2,5 GHz
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the electric breastpump Swing,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1,2√P
d = 1,2√P 80 MHz to 800 MHz
d = 2,3√P 800 MHz to 2,5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m). Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
b