Data Management; Complete A Fat; Disable Or Enable The Trigger Signal - Hach Event Monitor Trigger System Manual Básico Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

3. Push Plant Library. The plant event library shows.
4. Select the plant event to erase.
5. Push Delete, then OK.

Data management

Save the data logs to a USB flash drive
If the data logs are not saved to an external device (i.e., SCADA system) in real-time, periodically
save them to a USB flash drive for archival. The data logs are saved as CSV (comma-separated
value) files.
1. Install a USB flash drive into the USB connector.
2. From the Main Menu, push MAINTENANCE.
3. Enter the user name and password, then push OK.
4. Push File Save, then OK.
5. In the Export File Prefix field, enter a descriptive filename for the logs files to be transferred.
6. To quickly transfer all the information in the data logs and remove previously saved data logs
from the USB flash drive, push Reset.
7. Select the checkboxes for the data logs to be transferred.
8. Select Export to USB.
9. To safely remove the USB flash drive after the transfer is completed, select Eject USB on Exit.
10. Select the date range to transfer.
Option
Start from last export
point
Specify date range
11. Push Export.
12. Remove the USB flash drive from the USB connector.

Complete a FAT

The Complete FAT button on the configuration screen is for manufacturer use only.
Maintenance
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired,
contact the manufacturer.

Disable or enable the trigger signal

To prevent the occurrence of false plant events, set the trigger signal to off before a maintenance
task starts.
1. From the Main Menu, push MAINTENANCE.
2. Enter the user name and password, then push OK.
Description
Transfer the data log information that has not previously been transferred to
the USB flash drive.
Transfer the data log information for a selected date range. Enter a start date
and end date.