Shows the license key, license expiration date, license days that remain and software version.
5. Shut down the monitor. Refer to
6. Set the power to off and then on at the power disconnect device.

Enter the site information

1. From the Main Menu, push CONFIGURATION.
2. Enter the user name and password, then push OK.
3. Push Site Info. The current settings show.
4. Push Set System Date Time.
Note: Before the date and time setting is changed, get a new software license key from technical support. The
old license key will no longer work. Enter the new license key when prompted.
5. Select the date.
6. Enter the time.
Note: The hours, minutes, seconds and AM or PM are selected and changed individually.
7. Push Save, then OK to restart the monitor. If an earlier date is selected, all data with a future
timestamp will be erased.
8. Select an option.
Option
Site Name
Site ID
Location
Frozen Sensor
Session Timeout Set the amount of time of no activity before the current user is logged out.
TOC Present
Disinfectant
9. Push Save, then OK.

Configure the communication settings

If the monitor is supplied with an input control device (i.e., sc1000 controller) and/or a data output
device (i.e., SCADA), configure the communication settings.
Note: A loss in communication can occur when a change is made to the input networks.
1. From the Main Menu, push CONFIGURATION.
2. Enter the user name and password, then push OK.
3. Push Digital Com.
4. In the Input and/or Output section, select the communication settings.
5. Configure the register mappings. Refer to Appendix B of the GuardianBlue documentation.
6. Push Save, then OK.
14 English
Shut down the monitor
Description
Enter a unique name for the monitor.
Enter a unique identification number for the monitor.
Enter the name of the installation location.
Set the maximum amount of time that the sensor signals can stay constant before the
sensor signal is identified by the monitor as frozen.
Note: The nominal value entered for the sensor is used to calculate the trigger signal
value instead of the sensor signal when the sensor signal becomes frozen.
Select if a TOC analyzer is connected to the monitor.
Select the type of disinfectant used in the water system.
on page 23.