1.3 Attention
Please check the packing before use to make sure the device
and accessories are totally in accordance with the packing list, or else
the device may have the possibility of working abnormally.
The device may be a little different from the picture of manual, please make
actual device as standard.
Before using, check the main unit and all accessories periodically to make
sure that there is no visible damage that may affect patient's safety and
performance. It is recommended that the device should be inspected once
a month at least. When there is obvious damage, stop using the device
before changing the damaged parts.
Necessary maintenance must be performed by qualified service engineers
ONLY. Users are not permitted to maintain it by themselves.
The device is designed for continuous operation and is 'ordinary'. If the
device gets wet, please stop operating it.
Keep the device away from dust, vibration, corrosive substances.
Electromagnetic Interference-Ensure that the environment in which the device
is operated is not subject to any sources of strong electromagnetic
interference, such as MRI scan, mobile telephones, etc. Keep them far away.
Since the friction of body or clothes will bring high voltage static, and interfere
the veracity of the tested data by the apparatus, it is avoided to be used
in the environment of high voltage static.
This product is calibrated before leaving factory. Please don't measure this
device with functional tester for the device's related information.
The device has the function of limits alarming, when the measured data
is beyond the highest or lowest limit, the device would start alarming
automatically on the premise of the alarming function is on.
The device has the function of alarming, this function can either be paused,
or closed (default setting) for good, this function could be turned on through
menu operation if you need. Please check the chapter 4.3 as a reference.
When it is carried from cold environment to warm or humid environment,
please do not use it immediately.
DO NOT operate keys on front panel with sharp materials.
When cleaning and sterilization, please refer to user manual related chapters
for details.
The disposal of scrap instrument and its accessories and packings (including
battery, plastic bags, foams and paper boxes) should follow the local laws
and regulations.
The device has normal useful life of three years since the first electrified use.
ENGLISH
26