Turning the standing mode function off:
Requirement: The knee joint is fully bent.
1
2
INFORMATION
After the joint's confirmation with the beep signal, check whether the joint is still in the 1st mode (see Section
"Switching Between 1st, 2nd and 3rd Mode").
Optimisation of the Damping Behaviour in the 1st Mode
CAUTION
Risk of falling during optimisation of the damping behaviour.This adaptation changes the damping
behaviour of the joint. In certain situations, this can cause you to fall.
You must be standing in a secure position during this procedure.
This functionality enables you to adapt the damping behaviour of the C-Leg slightly during everyday use (e.g.
when getting used to the prosthesis or for a changed gait pattern). For safety purposes, the damping behaviour
can only be changed slightly with the remote control.
C‑Leg
1. Activation of the remote control: Press key 3 and kep it pressed (Fig. 1) until
the joint confirms the activation with a vibration signal.
2. Turning the standing mode function off: Within 3 seconds after activation of
the remote control simultaneously press the keys 1 and 2 (Fig. 2) until the joint
confirms turning off of the standing mode with one short beep signal.