Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM –
for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING
portable and mobile RF communications equipment and the ECG 300G
The ECG 300G is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the ECG300G can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the ECG 300G as recommended below, according
to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Manuale Utente
Recommended separation distances between
Separation distance according to frequency of transmitter
0.12
0.37
1.17
3.69
11.67
34
(m)
0.12
0.37
1.17
3.69
11.67
0.23
0.74
2.33
7.38
23.33