Installing the floor plate (Continued)
EN
6.
Using four 3/8 in. diameter socket head cap screws (F), anchor the floor plate into the recessed opening with the
counter-bored holes facing up (G)
vehicle.
A steel backing plate (or equivalent) is required for added support (not supplied).
One set of additional floor plates is supplied with the Model 6377 fastener to allow alternate positioning
anywhere on the patient compartment floor. Cot space requirements are the same as shown in
page
1-14. Repeat the above procedure to install the alternate position floor plates.
Figure 1-11: Floor plate anchor and locating points
7.
Position the antler over the floor plates and fasten the antler (A) with the red knobs (B)
Notes
Use the additional mounting holes in the antler bracket to position the antler for the Model 6371 wall mounted
cot fastener system.
The red knobs that fasten the antler assembly to the patient compartment floor plate are not compatible with
the universal floor plate. See the Universal Floor Plate Installation Instructions for floor plate installation.
8.
Position the rail support bracket (C) over the floor mounting plate (D)
9.
Insert the socket head screw into the keyhole and slide the assembly forward until the hole in the bracket lines up
with the tapped hole in the floor plate.
10. Fasten securely by tightening one of the red knobs (E)
11. Position the rail assembly (F) on the support bracket (G)
1-16
Installation
(Figure 1-11 on page
1-16). The keyhole should face toward the front of the
Figure 1-12: Position and fasten antler over the floor
(Figure 1-13 on page
(Figure 1-14 on page
6370-109-005 REV C
B
A
B
plates
(Figure 1-12 on page
(Figure 1-13 on page
1-17).
1-17).
1-17).
Figure 1-8 on
1-16).
www.stryker.com