14.5
Smart TV browser
The wrong time and the wrong date are displayed in the
graphical user interface.
Fig. 37:
Failure indication 'Invalid time format'
This failure indication has no effect on the result of
measurement.
i
Smart TVs with an Internet connection are not able
to connect to a time server to set the time and date
to the local time.
"
Connect Smart TV to the Internet. The date and time
are automatically set to the correct values.
14.6
Display problems with PC / smart
device browser
If the display in the browser is not correct, the
TTM 2104/2204 is not compatible with the operating system
being used, the browser being used or a combination of
both.
"
Load an alternative browser from the app store and
test it with TTM 2104/2204.
14.7
Incorrect font sizes
Fig. 38:
Incorrect font sizes
If the font sizes displayed in the browser cannot be
read, the "Zoom" setting is wrong.
"
Set the "Zoom" property to "100%" in the browser's
settings.
Robert Bosch GmbH
Print | TTM 2104/2204 | 47
15. Print
The TTM 2104/2204 software has an option to automatically
print a test protocol on completion of a measurement.
A printer dialog that must be confirmed by the user
appears in all browsers first.
In some browsers, it is possible to suppress this printer
dialog.
15.1
Firefox
¶
Download and install the add-on "Seamless Print"
from the website
https://addons.mozilla.org/enus/firefox/
.
addon/seamless-print/
Fig. 39:
Dialog for Seamless Print
"
Using the icon in the toolbar, you can easily switch
between display and hide the printer dialog.
en
en
|
1 691 206 007
2017-06-19