Pediatric Reduced Energy Defibrillation Electrodes Instructions for Use
Initial shock: Ultra Low Energy (150 VE) – All values are typical
Patient's Impedance
(ohms)
25
50
75
100
125
Initial shock: Low Energy (200 VE) – All values are typical
Patient's Impedance
(ohms)
25
50
75
100
125
Second and subsequent shocks: Ultra Low Energy (150 VE) – All values are typical
Patient's Impedance
(ohms)
25
50
75
100
125
Second and subsequent shocks: Low Energy (200 VE) – All values are typical
Patient's Impedance
(ohms)
25
50
75
100
125
Second and subsequent shocks: High Energy (300 VE) – All values are typical
Patient's Impedance
(ohms)
25
50
75
100
125
6
Phase 1
Voltage
Duration
(volts)
(ms)
370
6.1
550
7.3
640
8.6
705
9.8
770
11.1
Phase 1
Voltage
Duration
(volts)
(ms)
430
6.1
630
7.3
745
8.6
790
9.8
855
11.1
Phase 1
Voltage
Duration
(volts)
(ms)
370
5.8
550
6.5
640
7.0
705
7.4
770
7.8
Phase 1
Voltage
Duration
(volts)
(ms)
430
5.8
630
6.5
745
7.0
790
7.4
855
7.8
Phase 1
Voltage
Duration
(volts)
(ms)
500
5.8
700
6.5
820
7.0
920
7.4
960
7.8
Phase 2
Voltage
Duration
(volts)
(ms)
258
3.2
366
3.2
417
3.2
442
3.2
453
3.2
Phase 2
Voltage
Duration
(volts)
(ms)
298
3.2
422
3.2
482
3.2
511
3.2
524
3.2
Phase 2
Voltage
Duration
(volts)
(ms)
270
3.2
390
3.2
470
3.2
510
3.2
545
3.2
Phase 2
Voltage
Duration
(volts)
(ms)
295
3.2
425
3.2
510
3.2
560
3.2
610
3.2
Phase 2
Voltage
Duration
(volts)
(ms)
380
3.2
520
3.2
620
3.2
680
3.2
720
3.2
Energy
(Joules)
31
36
37
36
35
Energy
(Joules)
42
50
51
49
47
Energy
(Joules)
31
35
34
32
29
Energy
(Joules)
41
47
46
43
39
Energy
(Joules)
56
63
62
58
53
112-2013-005 F