5.2 Total station
5.2.1 Switching the total station on
NOTE
If the total station cannot be switched on, check that the batteries have been inserted correctly.
If the batteries have been inserted correctly and the total station still cannot be switched on, have it checked by Hilti
Service.
5.2.2 Setting up the total station
When starting up the total station the circular bubble level displayed on the screen must be centered by turning the
tripod footscrews.
NOTE
The total station is equipped with a laser plummet. The laser plummet can be used to set up the total station over a
point marked on the ground or floor.

5.3 Further information

Further information can be found at:
POS 150 (http://qr.hilti.com/r51292)
POS 180 (http://qr.hilti.com/r51294)

6 Calibration and adjustment

6.1 Calibration review

The tool is correctly adjusted when supplied. The values to which the tool is set may change over time or due to
temperature fluctuations, transport or aging. The tool therefore incorporates a feature that allows the settings to be
checked and, if necessary, corrected by carrying out in-the-field re-calibration. This is done by setting up the tool
securely on a tripod of good quality and targeting an easily visible, clearly discernible prism within ±3 degrees of
horizontal at a distance of approx. 50 - 70 m.
Switch the tool on or off.
Laser plummet on / off
Shift focus downwards, rolling.
Confirm the selected display.