Problem
... the display Err:12: appears?
... the display Err:16 appears?
... the display Err:32 appears?
8.2 Data transmission (models with a WiFi interface)
Problem
... the status LED does not come on?
... only the "SYS" item is visible in the "rF"
menu?
... after calling up the menu, the "rF" item is
not displayed?
... the display noChG appears when the WiFi
interface is activated?
... the status LED is illuminated red?
... the display Id:Err appears?
... Enter is pressed and the display Err:71
appears?
64 •
The scale has been switched on with too high a load.
- Remove the load from the scale
- Restart the scale
The scale was caused to oscillate and was unable to determine the zero
point.
- Restart the scale
An internal communication error has occurred.
- Restart the scale
- Inform seca Service
The status LED is faulty.
- Inform seca Service
The WiFi interface is deactivated.
- Activate WiFi interface
The scale's WiFi interface is faulty.
- Inform seca Service
Scale in battery mode, On option selected in rF\SYS menu
- Select PErM option
- Use power supply unit (recommended)
• ID
found in hospital information system (HIS) or in seca software (see
not
fault
).
Id:Err
• Device has
saved measured results to the clipboard.
not
- Repeat measurement
• Measured results have
- Deactivate BMIF function (see "Using the differential value function
(bMIF)" on page 57)
- Repeat measurement
- Check WiFi connection
The ID of the user or the patient has not been found in the HIS or in the
seca software.
- Scan the barcodes for the user and patient again
- Inform seca Service
No data transmission possible, WiFi interface is deactivated.
- Activate WiFi interface (see "Setting up the WiFi connection" on page 51)
Cause/solution
Cause/solution
been sent to an HIS or to seca software.
not