Changing The Fuse - Coltene BioSonic UC150 Guía De Uso

Ocultar thumbs Ver también para BioSonic UC150:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 104
12. Maintenance
Cleaning
All components should be cleaned each day by wiping with a soft cloth moistened with a
commercially available mineral deposit remover followed by cleaning with isopropyl alcohol or a
mild disinfectant. It is recommended to drain and dry the tank at the end of each work day.
Do not use petroleum based solvents, iodophors or phenolic based products.
Iodophors and phenolics can stain the surface of the unit. Clean up all liquid spills
immediately. Rust deposits on tank can be cleaned/ removed with a commercially
CAUTION!
available stainless steel cleaner.
Stainless Steel Cleaning
 Allowable:
• Wash with a clean cloth or rag soaked with warm water and mild soap or liquid dish detergent.
• Always clean in the direction of the original polish lines.
• Always rinse well with clear water (2 or 3 times) after cleaning. Wipe dry completely.
• Specialized household stainless steel cleaner may also be used.
 Avoid:
• Do not let dirt accumulate.
• Steel or stainless steel wool or scrapers to remove stubborn debris.
• Harsh or abrasive cleaners.
• Chemicals harmful to stainless steel not limited to but including products that contain
chloride, fluoride, iodide or bromide.
• Cleaners that contain bleach.
In case of questions regarding compatibility of decontamination or cleaning agents please contact
the manufacturer or your supplier.
Inspecting the Power Cord Periodically
Check the power cord for damage and see that the connectors at both ends of the cord are fully
seated.

Changing the Fuse

• Remove the power cord from the wall outlet first and then from the power cord receptacle on
the rear of the unit.
13

Publicidad

Tabla de contenido
loading

Tabla de contenido