• The fuse compartment is located directly above to the power cord receptacle. To access, take a
small, flat-headed screwdriver to the tabs on the fuse holder. Apply slight pressure to release
(Figure
the fuse holder
• Gently remove the fuses from the fuse holder.
(Figure
• Examine the fuses
current. If fuse appears to be in good condition, return the unit to your dealer or Coltène/
Whaledent. This is an indication that the unit is defective.
• Plug the power cord into the power cord receptacle on the rear of the unit. Then plug the power
cord into the wall receptacle.
NOTE: The fuse is a protection device designed to limit the amount of current being drawn or
supplied to the unit. If a fuse is blown, this can mean that there was a surge of current to the unit
due to an external power surge such as lightning, or more likely due to an internal component
failure of the cleaner.
Firmware Update Process

Checking the Firmware Version

• Press the "Setup" icon from the main menu, bottom left icon on the display
• Press the "Information" icon from the setup menu, far right icon on the display
• The unit serial number and firmware version will be displayed
• Press the "Check Mark" icon then the "Home" icon to return to the main menu
Firmware Update Process
• Place the most current firmware update folder "coltene-update" on a blank USB stick
• With the BioSonic UC150 displaying the main menu, place the USB stick in the display's USB port
• Wait a minute and the software update screen will appear on the display noting the current
firmware version on the BioSonic UC150 and the current version on the USB stick
• Press the "Update" icon to start the software update process
• Do not touch the unit or the USB stick during the update process until the unit reboots and the
main menu is displayed
• The firmware update process is complete and the USB stick can be removed
*see last page for figures
14
9*).
10*). If blown, replace with a fuse that matches the unit's voltage and