INDICATIONS FOR USE

The Abbott FreeGo enteral feeding pump is intended for hospital, acute care, long-
term care, and home use. It is designed for both stationary use and ambulatory
use. The pump is also designed to function safely on a commercial aeroplane.
The Abbott FreeGo pump can be used for adult and paediatric patients,
provided the patients can tolerate a feeding range within the pump operational
specifications.
• The flow rate range is 1 to 400 mL/hr, in 1 mL/hr increments.
• The flow rate accuracy is +/- 5%, under recommended head height (see
PUMP ACCURACY TESTING).
• Pump operates against 103 kPa to 124 kPa nominal back pressure before
occlusion alarm.
If these specifications are not appropriate for a given patient, the Abbott
FreeGo pump should not be used.
Page 36
44838 Rev. J