DIAL /
BUTTON
Page 56
FUNCTION
The Prime Button is used to remove air from the feeding set. This button
is functional only in the FLUJO (SET RATE) and DOSIS (SET DOSE) dial
positions. While priming, be absolutely certain that the feeding set is not
connected to your enteral feeding tube. See INSTRUCTIONS FOR USE.
The pump will begin auto priming if this button is depressed when the pump's
sensors detect air in the feeding set and the pump dial has not yet been set to
MARCHA (RUN) after one of the following has occurred:
•
The door has opened and closed, or
•
The pump dial is turned from OFF/CARGA (OFF/CHRG) to any other
position.
Otherwise, the pump will begin incremental priming when this button is
depressed.
The Options Button is used to change the audio volume of the alarms, as
well as the brightness and timing of the LCD display. Press the button until
the option you wish to change is displayed, then use the Up and Down Arrow
Buttons to choose your preference:
NIVEL SONIDO (AUDIO LEVEL)
- BAJO (LOW)
- ALTO (HIGH)
BRILLO (BRIGHTNESS)
- BAJA (LOW)
- MED (MED)
- ALTO (HIGH)
LUZ (LITE- controls when display is on)
- 10 SEG (10 SEC ON)
- SIEMPRE (ALWAYS ON)
NOTE: "SIEMPRE" ("ALWAYS ON") option is only available when pump is
plugged in to live AC/Mains, in order to conserve battery charge.
COMM (for "communication", controls which Accessory Cable is enabled for
communication)
- NC
- DATA
NOTE: See INSTRUCTIONS FOR USE for complete details
44838 Rev. J