ALARM AND WARNING MESSAGES

PRIMARY SECONDARY
DISPLAYS
OBST
NO ENTRA ALIM
(OCCL INTO PUMP)
OBST
NO SALE ALIM
(OCCL OUT OF PUMP)
ATN
BOMBA EN PAUSA
(ATTN PUMP ON HOLD)
BAT
CARGA BAJA
(BATT LOW BATTERY)
BAT
AGOTADA
(BATT BATT EMPTY)
ATN
PUERTA ABIERTA
(DOOR OPEN)
CONDITION OR
PROBLEM
Flow has stopped
due to occlusion
between the formula
container and the
pump.
Flow has stopped
due to occlusion
between the pump
and the patient.
Hold time has
elapsed.
Pump battery is
nearly empty. At least
20 minutes of battery
capacity is remaining
from time of first
alarm.
Pump battery is
depleted. Pump has
stopped feeding.
The door is open
while feeding or
priming the feeding
set.
44838 Rev. J
EXPLANATION OR CORRECTIVE
ACTION
Turn dial to PAUSA (HOLD)
Remove feeding set from pump.
Clear obstruction; check for kinked
tubing or empty container. Reinstall
feeding set.
If this alarm appeared during priming,
attempt to prime the feeding set
again, then connect the feeding
adapter to your enteral feeding tube
Turn dial to MARCHA (RUN) to begin
feeding
Turn dial to MARCHA (RUN) to begin
feeding
Connect the pump to live AC/Mains
power.
Connect the pump to live AC/Mains
power. The display will indicate
"CARG POR FAVOR ESPERAR"
("CHRG PLEASE WAIT")
Turn dial to PAUSA (HOLD)
Close the door
Turn dial to MARCHA (RUN) to
begin feeding, or attempt to prime the
feeding set again
Page 49