» INCREMENTALLY PRIME THE SET: There may still be a small section
of air in the feeding set. Press and hold the Prime Button. Priming will
begin after two (2) seconds. Release Prime Button when fluid has
reached desired level.
NOTE: See
in PUMP DIAL AND BUTTON REFERENCE for more information on
priming.
TO START PUMP:
1. Turn pump dial to FLUJO (SET RATE). Then select your prescribed flow
rate by pressing the Up and Down Arrow Buttons while observing the entry
on the display. Keeping an Arrow Button depressed will cause the scroll rate
to increase.
NOTE: Any time the dial is turned from OFF/CARGA (OFF/CHRG) to any
other dial position (i.e. the pump is turned on), the pump will initiate system
Self-Check Procedure. During this procedure, user should verify that all
segments on the LCD display activate and deactivate, then the pump's
software version and serial number (for example, V1:00 AF11001000) are
displayed. Simultaneously, verify that the audio beeps at high volume, then
low, then high again. If the pump has a software version later than V1:01,
then the pump will also display either NC ENABLED or DATA ENABLED,
at the end of the sequence described above. Do not use the pump if the
operation is not exactly as described above.
2. Turn the dial to DOSIS (SET DOSE) and programme your prescribed dose
using the Up and Down Arrow Buttons. Keeping an Arrow Button depressed
will cause the scroll rate to increase.
NOTE: Pump can be programmed to feed until the container is empty
(DOSIS (SET DOSE) at INF DOSIS (infinite dose); see PUMP DIAL AND
BUTTON REFERENCE). In this case, feeding will continue until the pump's
sensors detect air in the feeding set. At that time, the pump will stop and
the display will read "NING ALIMENTO" ("FEED EMPTY").
NOTE: Be sure that the feeding volume has been cleared from the pump's
memory prior to beginning a new feeding (for instruction, see VOL CERO
(CLEAR VOL) at step 6, below).
3. Connect the Abbott FreeGo feeding set securely to the enteral feeding tube.
4. Turn dial to MARCHA (RUN) to start feeding.
NOTE: If alarm sounds, turn pump dial to PAUSA (HOLD). Correct alarm
condition that has been indicated on the LCD display, then return dial to
MARCHA (RUN) to restart feeding. For Alarm and Warning information,
see ALARM AND WARNING MESSAGES.