Page 54
FUNCTION
Primary purpose is to set the dose (to limit the total amount of fluid
volume) delivered during an entire feeding.
To program feeding dose, use Up and Down Arrow Buttons. Keeping
an Arrow Button depressed will make the numerals scroll. The scrolling
rate will increase (depending on length of time Arrow Button is
depressed).
If you wish to feed until the container is empty, set the dose to INF
DOSIS (infinite dose). To do this, make sure the pump dial is in the
DOSIS (SET DOSE) position then depress the Down Arrow Button
until INF DOSIS appears on the display.
Auto Prime and Incremental Prime features are functional in this dial
position.
To set the lock-out feature, simultaneously press both Up and Down
Arrow Buttons. To deactivate lock-out feature, repeat procedure.
If lock-out feature is not set, the display will read "FIJE DOSIS" ("SET
DOSE"), along with the current dose setting (from 1 to 9999 mL, or
INF).
If lock-out feature is set, the display will read "BOMBA BLOQUEADA"
("PUMP LOCKED").
Primary purpose is to initiate and continue a feeding.
This is the only dial setting that operates a feeding.
As long as dial is kept in this position, feeding will continue until:
»
The volume delivered equals the programmed dose volume
(when VOL ADM equals DOSIS),
or...
»
An alarm condition has been detected.
Display will show the programmed rate.
Display will scroll 4 drops in bottom-right corner.
It is normal for rotor movement to be intermittent (e.g. starts, stops,
starts, etc.) while in MARCHA (RUN) mode. This intermittent
movement regulates the rate of flow, conserves battery charge, and
does not affect the accuracy of the pump.