CAUTION: Be certain to verify functionality at the nurse call station prior to
connecting the feeding set to the patient.
To verify functionality of the Nurse Call Accessory Cable (S468), turn the pump dial
to RUN. Then open the pump door, which will trigger a DOOR OPEN alarm. Verify
the nurse station is receiving this alarm. Press the nurse call button and verify the
nurse station is receiving this alarm.
CAUTION: The Nurse Call Accessory Cable functionality should be verified
periodically.
TROUBLESHOOTING:
If the NC function is not working properly:
• Check to see that all ends of the accessory cable and charger are securely
connected.
• Check the accessory cable for damage.
• If the cable is properly connected and not damaged, return the pump and
cable for service.