USING THE DATA DOWNLOAD ACCESSORY CABLE (S412):
CAUTION: The pump COMM (communication output) setting must be configured
to DATA (for Data Download Accessory Cable).
NOTE: Pumps with software version V1:01 are not configured for use with the Data
Download Accessory Cable (S412). To check the software version of your pump,
from the OFF/CHRG position, turn the pump dial to SET RATE. During the system
Self-Check Procedure the pump display will show the software version. If your
pump has software version V1:01, contact your local Abbott representative to have
the pump upgraded.
To configure your pump for use with the Data Download Accessory Cable:
Press the Options Button
Then use the Up and Down Arrow Buttons until DATA is displayed.
Pump must be plugged into live AC/Mains power in order for Data Download Accessory
Cable to work. Refer to the drawing below for cable and charger connections.
To verify functionality of the Data Download Accessory Cable (S412), turn the pump
dial to RUN. Then open the pump door, which will trigger a DOOR OPEN alarm.
Verify the alarm is transmitted from the Data Download Accessory Cable.
TROUBLESHOOTING:
If the DATA function is not working properly:
• Check to see that all ends of the accessory cable and charger are securely
connected.
• Check the accessory cable for damage.
• If the cable is properly connected and not damaged, return the pump and
cable for service.
Page 46
until COMM is displayed on the pump display.
44838 Rev. J