Abbott FreeGo Manual De Uso página 63

Bomba para alimentación enteral
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

STORAGE
Storage Temperature:
Storage Humidity:
Atmospheric pressure:
POWER
Voltage:
Power consumption:
Type of protection fuse:
Operating voltage fuse:
Current opening fuse:
Type of protection against
electric shock:
Degree of protection against
electric shock:
Degree of protection against
harmful ingress of water:
Degree of safety in the
presence of a flammable
anesthetic mixture:
BATTERY
Type:
Charge Life:
Charge Time:
When the alarm message "BATT LOW BATTERY" first appears, there is at least
20 minutes of battery capacity remaining for pump operation.
• Pumps must be charged for a minimum of 6 hours prior to use to ensure
optimum performance.
• The FreeGo pump battery may be fully discharged when received and
may require approximately 2 minutes of AC power prior to illumination of the
AC/Mains LED indicator.
-20° to +65 ° C
10% to 95% non-condensing
50 kPa to 106 kPa
Charger input 100-240 V; 50-60- Hz; output 5 V DC 2.4A
320 mW typical (Run Mode, 125 ml/hr)
Single blow fast acting
125 V
2 A
Galvanic isolation by transformer and opto-coupler
Class II, double insulated
IP25
Possible explosion hazard - Do not use
Lithium Ion, 3.7 volt
≥ 24 hours, at 125 mL/hr
≤ 6 hours
44838 Rev. J
Page 61

Hide quick links:

Publicidad

Capítulos

Tabla de contenido
loading