Descargar Imprimir esta página

Installation - Agilent Technologies Turbo-V 81-M Serie Manual De Instrucciones

Publicidad

Installation

Do not remove the adhesive and protective cap before connecting the turbopump
CAUTION!
to the system.
Figure 2
Do not install or use the pump in an environment exposed to
atmospheric agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk.
During operation, the following environmental conditions must be
respected:
In the presence of magnetic fields the pump must be protected using
a ferromagnetic shield. See the appendix "Technical Information" for
detailed information. The Turbo-V 81-M pumps must be used in
conjunction with one of the suitable Agilent controller and they must
be connected to a primary pump (see "Technical Information").
Turbo-V 81-M User Manual / 87-900-983-01 (F)
maximum pressure: 2 bar above atmospheric pressure
temperature: from +5 °C to +35 °C (see the diagram pres-sure-
temperature in the appendix "Technical Information")
relative humidity: 0 – 95 % (non-condensing)
Instructions for Use

Installation

193/237
16

Publicidad

loading

Este manual también es adecuado para:

969-8901969-8902969-8903969-8904