Height adjustment help message
If the Ready LED keeps flashing, Help LED (19) for
height adjustment on Control Panel A is illuminated
and there are no dots under the three digits on the height
adjustment display (FH position is correct), this can
mean:
The unit is too high or too low for the teleradiograph
scan (for example, after a previously taken panoramic
exposure, since the height-adjustment range of the sys-
tem is limited for the teleradiographic function).
Then move out of the too high or too low range using the
↑ or ↓ button on Control Panel A until Help LED (19)
extinguishes.
This will happen while the unit is moving.
ATTENTION
Under no circumstances must the patient then be in the
head support!
If the Ready LED still continues to flash, the cause can
be tracked down by calling up help messages H3... or
H4... and the necessary steps taken to eliminate the
problem.
Consult the list of help messages H3.../H4... on the
inside of the doors.
i
NOTE
Press button T and then the exposure button to initiate a
test run of the teleradiograph scan without radiation.
Only release the exposure button when the unit has
stopped moving!