You can clean and disinfect the exterior surface of the PowerPark stand module using a recommended
cleaner or disinfectant. For the most recent list of approved products, see Cleaning and Disinfecting FUJIFILM
SonoSite Products, available at www.sonosite.com.
Caution
1 Turn off the ultrasound system and any accessory equipment. Disconnect the stand from the PowerPark
dock or unplug the power cord.
2 Clean the stand module surfaces to remove any debris or bodily fluids. Use the following procedure:
a Use either a pre-moistened wipe or a soft cloth dampened with cleaner or disinfectant.
b Remove all gel, debris, and bodily fluids from the PowerPark stand module.
3 With a new wipe, clean all parts of the PowerPark stand module by wiping from the clean areas to the
soiled areas. This method helps to avoid cross-contamination.
Follow the chemical manufacturer's instructions regarding temperature and wet contact time. Monitor the
PowerPark parts for wet appearance, and reapply with a new wipe if no longer wet.
4 Allow the PowerPark stand module to air dry in a clean, well-ventilated space.
To clean the PowerPark dock
Clean the PowerPark dock periodically, or whenever it appears dirty or comes into contact with biomatter or
bodily fluids.
Caution
1 Undock the stand, and disconnect the PowerPark dock from the AC power.
2 Clean the outside surface of the PowerPark dock to remove any debris or bodily fluids. Use either a pre-
moistened wipe or a soft cloth dampened with cleaner or disinfectant.
3 With a new wipe, clean all parts of the PowerPark dock by wiping from the clean areas to the soiled areas.
This method helps avoid cross-contamination.
Follow the chemical manufacturer's instructions regarding temperature and wet contact time. Monitor the
PowerPark parts for wet appearance and reapply with a new wipe if they are no longer wet.
4 Allow the PowerPark dock to air dry in a clean, well-ventilated space.
8
Do not allow cleaning solution or disinfectant to contact the PowerPark electrical
connectors.
Do not allow cleaning solution or disinfectant to come into contact with the
PowerPark electrical connectors.