Safety
For maximum safety, observe all of the warnings and cautions in this section.
Electrical safety
Table 2: Electrical safety classification
Class I equipment
Non AP/APG
WARNINGS
Safety
The PowerPark is classified as Class I equipment.
The PowerPark is not suitable for use in the presence of
flammable anesthetics.
To avoid the risk of a shock hazard:
Do not plug the power cord of the PowerPark dock or PowerPark stand
module into a multiple portable socket outlet (MPSO) or use an extension
cord.
Make sure the AC power cord is disconnected from the PowerPark stand
module whenever the stand is docked.
To avoid the risk of electrical shock and possible fire:
Use only properly grounded equipment. Shock hazards exist if the power
supply is not properly grounded. Grounding reliability can only be
achieved when the equipment is connected to a receptacle marked
"Hospital Only," "Hospital Grade," or the equivalent. The grounding wire
must not be removed or circumvented.
Use only accessories and peripherals recommended by FUJIFILM
SonoSite, including the power supply. Use the PowerPark only within the
voltage and current limits specified on its ratings label.
Inspect the power supply, plugs, cables, and power cords on a regular
basis for damage.
Do not place the PowerPark dock in a wet environment.
To avoid system malfunction and the risk of electrical shock, follow the
PowerPark assembly instructions.
9