Ejemplos De Instalaciones - SICK FLOWSIC500 Instrucciones De Servicio

Caudalímetro ultrasónico de gas con conversión opcional del volumen
Ocultar thumbs Ver también para FLOWSIC500:
Tabla de contenido

Publicidad

Anexo

Ejemplos de instalaciones

9 .6
Fig. 72
Funcionamiento con batería
FLOWSIC500 with LF output connected to electronic volume corrector
2 1
3 4
M12connector
B-coded
1 n.c.
2 n.c.
3 - DO 1 (passive)
4 + DO 1 (passive)
FLOWSIC500 · Instrucciones de servicio · 8015394/13GE/V 3-3/2019-10 · © SICK Engineering GmbH
(both battery powered and intrinsically safe)
Hazardous area
EVC
battery powered
intrinsically safe
[Ex ia]
impulse
-
error
-
2 1
3 4
M12 connector
A-coded
1 + DO 2 (passive)
2 - DO 2 (passive)
3 - DO 3 (passive)
4 + DO 3 (passive)
FLOWSIC500
battery powered
intrinsically safe
[Ex ia]
M12 connector A-coded
1 + DO 2 (passive)
2 - DO 2 (passive)
4 + DO 3 (passive)
3 - DO 3 (passive)
For safety and entry parameters see
control drawing 9215965 for the USA and Canada
and control drawing 9215966 for the EU.
In the USA install in accordance with the NEC (ANSI/NFPA70)
In Canada install in accordance with CEC part 1,
in the EU in accordance with EN 60079-14.
National regulations must be observed!
WARNING!
Incorrect cabling can cause the FLOWSIC500 to fail!
See Operating Instructions for further details!
PE
PE
143