3.2 Interface parameters

LocalCAN: No transmission parameters can be set
Contact
close <1 sec: On and off switching of the selected function
close >2 sec: Discharge
close >10 sec: Error message "Short circuit at contact socket "
RS232
9600 baud, no parity, 8 data bits, switchable to
2400 baud, even parity, 7 data bits.
software handshake Xon/Xoff switchable to Hardware handshake DSR/DTR.
(See section "Settings".)

3.3 SICS command set of the LV11

The following SICS commands (in boldface) are recognized and acknowledged (normal font) at the
CAN and the RS interface if any type of device is attached to these interfaces. (Char "_" = space). All
commands must be closed with CR and LF.
LV01
Reset (LV11 returns to normal mode, all parameters are reset)
LV01_A
Reset executed (also after switching on)
LV02_X
Set intensity X. X may be between 1 and 4.
LV02_X_A
Command "Set intensity" executed
LV02_L
Command "Set intensity" understood, parameter faulty
LV02_I
Command "Set intensity" not executable as previous process has not been completed.
LV03
Vibrate until passage through the light barrier has been detected (LV11 in
vibrate mode)
LV03_B
Command understood (vibratory action starts)
LV03_A
Command executed (light barrier passage, vibratory action has stopped)
LV03_I
Command not executable as previous process has not been completed or command
ended by timeout of 90 seconds.
LV04
Vibrate until a stop or reset command is sent or the stop key has been pressed
LV04_B
Command (vibratory action starts)
LV04_I
Command not executable as previous process has not been completed.
LV05
Discharge (LV11 in discharge mode)
LV05_B
Command understood (discharge started)
LV05_A
Discharge ended (90 seconds after last passage through light barrier)
LV05_I
Command not executable as previous process has not been completed.
LV06
Stop (LV11 returns to normal mode)
LV06_A
Stop executed