3.1.3

Wiring Temperature Sensor to Transmitter

Wiring Temperature Sensor Board to Transmitter
Electrical Shock Hazard
Contact with exposed wiring will lead to fire, electric shock, or serious personal injury.
Set transmitter and instrumentation power to OFF when inserting or removing the Analog
Input Module, or when making connections to TB1, TB2, TB3 and TB4.
1. Disconnect power to the transmitter.
2. Open transmitter top cover by releasing the cover latch.
3. Loosen the captive screw securing the Access Cover and remove Access Cover.
4. Using a flat-blade screwdriver, remove four screws securing the I/O Module board.
FUT1010 IP65 NEMA 4X & IP66 NEMA 7 Liquid Quick Start
Operating Instructions, 1/2013, A5E02639182-AC
DANGER
Remove board and set it aside.