Integra
Padgett
®
Electric Dermatome Models B and PI
How to Use the Electric Dermatome Models B and PI
(continued)
The desired thickness of the graft is
predetermined by the operator and set
by using the pointer on the thickness
guide scale. Secure the thickness knob
pointer position by tightening the knob's
thumbscrew. It is important to ensure that
the knob's thumbscrew has been loosened
prior to moving the knob pointer. The actual
thickness of the harvested tissue is highly
dependent upon the operator technique and
the condition of the tissue being harvested.
The thickness guide scale provides both inch
and millimeter values.
Much of the weight of the motor should
be carried by the hand that holds the
dermatome, to maintain a reasonably light
pressure between the cutting edge and the
patient's body. The instrument should be
held at a 25 to 45 degree angle.
The cut graft automatically folds itself into
the "pocket" of the dermatome head. To sever
the graft, simply lift the dermatome up and
away from the donor site. As per Association
of Perioperative Registered Nurses (AORN®)
recommended practice, the dermatome's
handpiece cord should be secured to the
sterile field with a non-perforating device.
17
®