Integra
Padgett
®
Electric Dermatome Models B and PI
APPENDIX (continued)
Recommended separation distances between portable and mobile RF
Rated Maximum
Output Power of
Transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
27
®
communications equipment and the equipment.
Separation distance according to frequency of transmitters in
meters
150 kHz - 80 MHz
d=1.2√P
0.12
0.37
1.17
3.69
11.67
80 MHz to 800MHz
d=1.2√P
0.12
0.37
1.17
3.69
11.67
800 MHz to 2.5 GHz
d=2.3√P
0.23
0.74
2.33
7.38
23.33