4 – 9 Irrigation nozzle attachment
X-SG20L and SG20 has three irrigation methods available depending on the tool and application: external, internal or
both� For details on installation, refer to the included handpiece operation manual�
4 – 10 Attaching the tube holder
Use the motor cord as a strain relief for the irrigation tube� It is easier to insert the motor cord first, then the irrigation
tube�
Motor cord
Irrigation tube
4 – 11 Check before treatment
Perform the following checks before using the product on a patient to confirm that there are no abnormalities�
If abnormal vibration, noise, or overheating is detected, stop using the product and contact your authorized NSK dealer�
1
Make sure each component is securely connected�
2
Make sure the control unit main power switch is in
the OFF position (
side), and then connect to a
commercial power socket�
3
Put the control unit main power switch in the ON
position ( side)�
4
Confirm compatibility as covered in "4 – 11 – 1
Compatibility check of internal spray nozzle/drill"�
● Touching the control panel of an implant micromotor may increase the risk of cross-infections� In order to
minimize such a risk, it is possible to use a barrier film in combination with Surgic Pro2� Do not use a film with
poor operability and/or visibility�
5
Run the motor for 1 minute and check the following�
– Is the bur or drill free from rattling, abnormal
vibration, or abnormal noise?
– Can irrigation be done and is the coolant solution
flow level appropriate?
– Does the motor light illuminate (optic motor only)?
6
Stop the motor and make sure there is no abnormal
heating of the motor and handpiece surface�
CAUTION
15